WAT4638 Display Image

Gonotec Osmomat 050 Colloid Osmometer with Consumables Starter Kit

Code: WAT4638 N1-1

The Osmomat 050 has been especially developed for use in intensive care units. It has been designed to be used on a 24 h/day 7 day/week basis and the system is always ready to collect data. The sys...


read more

Your Price
€10,901.80 EACH
€13,409.21 inc. VAT

The Osmomat 050 has been especially developed for use in intensive care units. It has been designed to be used on a 24 h/day 7 day/week basis and the system is always ready to collect data. The system performs all cleaning and control operations during the stand-by period so that the user can perform a measurement at any time. After performing a measurement, the Osmomat 050 is automatically cleaned so the next measurement can be performed as required.

Due to the high degree of automation of the measuring and control functions of the Osmomat 050, the operator interaction is reduced to a minimum. This ensures that the instrument is always ready for measurement and the instrument is especially suitable for application in intensive care units. Typical applications for the Osmomat 050 include the measurement of the colloid-osmotic pressure with blood loss, hypalbumenia, albumin therapy for the prevention of lung edemas and all diseases resulting in a change of the colloid-osmotic pressure.

  • High-performance measuring cell
  • Minimum sample volume
  • Simple injection of the sample solution (plasma or serum, whole blood not recommended)
  • Measuring time ﹤1 minute
  • Automatic zero setting
  • User selection of the pressure dimension (mm Hg), (cm water column) or (k-Pascal), respectively
  • Display of the first result and storage of a second value serve as a control of the equilibration of the system
  • Automatic rinsing of the measuring cell after each measurement
  • Simple calibration with hydrostatic pressure differential
  • Automatic stand-by function over longer periods of time

Measurement volume﹤10µL
Sample size﹥100µL
Measurement time1-3 minutes depending on membrane type
Measurement format99.9 mm Hg, 99.9 k-Pascal
Reproducibility﹤±0.5 mm Hg
Significant figures3 places
Zero point determinationAutomatic
Membrane materialTwo layer membrane, depending on type retention of molecular weights > 10,000 - 30,000 Dalton
Recorder output2 mv/digit, ±10%, can be set for recorder compensation with 1 volt output
CalibrationDirect pressure measurement via difference in hydrostatic pressure
Automation1. Standalone wash of the upper half of the cell is available to prepare the cell; 2. Storage of the result; 3. Storage of an additional value over a specified time period after the first measurement